Loading...
Seca
_
Sheet
2(2)
_
Description
of
faults
Description
of
the
fault
Possible
cause
Remedy
Er
:X:
14
Error
during
calculation
of
the
kg
value
Incorrect
coefficient
in
the
read-only
memory
of
the
module;
recalibrate
the
scales
Er
:x:
15
Error
during
recalibration
Reference
weight
is
outside
the
permissible
range
or
operator
error
Er
:x:
2
1
Error
while
measuring
the
battery
voltage
Battery
voltage
is
far
too
low
(<
2.5
V)
or
electronics
are
defective
->
Replace
the
electronics
Er
:x:2
2
Error
while
measuring
the
temperature
Replace
the
electronics
Er
:x:3
2
Error:
command
buffer
overflow
Check
the
cabling
of
the
bus,
eliminate
any
EMC
malfunction,
briefly
(60
sec.)
de-energize
the
electronics
E
r
:x:4
0
No
EEProm
data
Recalibrate
the
scale
or
write
the
correct
EEProm
data
by
service
program
Er
:x:4
1
Incorrect
EEProm
access
by
the
slave
Check
the
slave;
disconnect
or
replace
the
slave
Er
:x:5
0
Other
errors
Error
occurs
directly
after
switching
on
->
wrong
checksum
in
the
ROM
->
Replace
the
electronics
Er
:x:5
1
Menu
error
Incorrect
use
of
a
menu
function
(wrong
data
e.g.),
try
it
again
07.07.99-
Meyer
30-34-00-588