Seko OPL Mini L - Programming the lock-out time; Calibration mode; Changing access code; Setup Trigger HI-LO

Seko OPL Mini L
2 pages
Print Icon
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPL Mini L
code 0000136939
rev. 1.2
2
3.1.4
Programming the lock
-
out time
-
Enter to programming mode (see par. 3.1.1).
-
Press the
key until the green display shows
3
”.
-
Press the PRG key. The red display’s first digit begin
s to blink.
-
Press the
key until the blinking digit displays the desired value.
-
Press the PRG key. The next digit blinks.
-
Press the
key until the blinking digit displays the desired value.
-
Press the PRG key. The next digit blinks.
-
Press the
key until
the blinking digit displays the desired value.
-
Press the PRG key. All of the display’s digits stop blinking.
-
To save the settings, while none of the display’s digits are blinking, scroll the menu using the
key until
Hld
appears on the
display and then
press PRG key. The pump returns to standby mode and
on
appears on the display.
3.1.5
Calibration mode
-
Enter to programming mode (see par. 3.1.1).
-
Press the
key until the display shows
5CAL
”.
-
Press the PRG key, CAL begins blinking.
-
While holding a measurin
g cup or flask under the outlet of the pump, press the
key and release when the pump starts
(display
increases the timer)
. Let the pump runs until desired amount of chemical is dispensed then press the
key again to stop. (display
shows the dosing time)
. The
OPL Mini
run time is now programmed. Repeat these steps if new volume is required.
-
Press the PRG key the display shows
5CAL
”.
-
To save the settings, scroll the menu using the
key until
Hld
appears on the display and then press PRG key. The pump
returns
to standby mode and
on
appears on the display.
3.1.6
Changing access code
-
Enter to programming mode (see par. 3.1.1).
-
Press the
key until the display shows
o+
old access code
”.
-
Press the PRG key the display shows
n+
old access code
and the red di
splay’s first digit begins to blink.
-
Press the
key until the blinking digit displays the desired value.
-
Press the PRG key. The next digit blinks.
-
Press the
key until the blinking digit displays the desired value.
-
Press the PRG key. The next digit blink
s.
-
Press the
key until the blinking digit displays the desired value.
-
Press the PRG key. All of the display’s digits stop blinking.
-
To save the settings, while none of the display’s digits are blinking, scroll the menu using the
key until
Hld
appears
on the
display and then press PRG key. The pump returns to standby mode and
on
appears on the display.
3.2
SETUP TRIGGER HI
-
LO
-
Enter to programming mode (see par. 3.1.1).
-
Press
the
key until the display shows
4 HI
or
4 LO
”.
-
Press the PRG key and
“HI”
or “LO” readout will blink
.
-
Press
▲ key
to select if a signal is
LO
(20÷240 VAC)
or
HI
(70÷240 VAC)
.
-
Press PRG key and the display stops blink
-
To save the settings, while none of the display’s digits are blinking, scroll the menu using the
key until
Hld
appears on the
display and then press PRG key.
The signal trigger has a filter of 5 seconds.
3.3
EXITING FROM PROGRAMMING
To exit programming and return to the main menu, scroll the menu using the
key until
ESC
appears on the display.
Now p
ress PRG key.
The pump returns to standby mode and
on
appears on the display.
None of the modified settings are saved.
3.4
PRIMING MODE
On the standby mode (the display shows the message: on) push the
key the pump starts dosing for one minute and the
display shows
Pri
”.
To stop the pump before the priming time is elapsed, press
▲ key to stop.
3.5
MOTOR STOPPING
In case the motor is excessively overloaded (for ex. a tubing breaks) the pump executes stop/restart motor cycles (the stop time increases at
every STOP: 0.5 sec .
10 sec.) to avoid the overheating of the same pump.
During the motor stop the display visualizes:
1Alr
4.0
DISPLAY
As soon as the pump is powered, the display shows
8888
for approximately 0.3 seconds, during which the pump ru
ns a check:
If the parameters cannot be read properly from EEPROM, the display shows
FAIL
and the pump remains on standby.
The default parameters can be reset using the
key (pressed for 4 seconds). The default parameters are:
Delay time = 005
Dosing time = 055
Lock
-
out time = 000
Trigger signal = HI
Calibration = CAL
Old password = 000
After the default parameters have been reset, these parameters can be mod
ified from the programming menu (see chap. 3).
5.0
ACCESSORY
Available
a remote control, the code
is
9900700097.