INSTRUCTION MANUAL CODE 80386 REV A FEBRUARY 2020 Page: 22
J.P. SELECTA s.a.u. Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España Tel 34 937 700 877 Fax 34 937 702 362
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
7.4 PROGRAMMING screen
Go to PROGRAMMING screen to:
Set the temperature proles.
Set the timed operation for a prole.
(For instance, you can set a time and a week day to start a prole. As:
start, prole n
er
1 on Sunday at 19h00’)
7.4.1Programmingatemperatureprole:
A temperature prole is a sequence of steps, also called “dells”.
To go from the end of a «dell» to the following «dell», it could be done in
two ways:
- With a «temperature ramp».
(go from 100ºC to 200ºC following a ramp at 2ºC per minute).
- As fast as possible, “--” indicator (maximum power).
Prole parameters input:
Select the prole number between 0 and 10 with .
Press on the parameter to be modied. (Temp, ramp, length ...)
Press on these icons: to enter the value.
Modify another parameter.
Special values:
If you enter value OFF in temperature (you have to press Increase key)
and the next steps are disabled and not executed.
If you enter INF in Step length, the oven remains permanently at this
temperature.
If you enter -- in the ramp, the oven makes the temperature rise at the
fastest speed possible.
Starting of temperature prole
To set the prole starting time, press on «—:—» (09:00 in the picture).
Enter time by pressing in hh:mm format.
Press on the days of the week when the prole should be run.
Accept the programming by pressing on .
Note: If you run a timed program, do not forget to press the RUN key to start
the program. During the timeout, it will indicate that it is in Standby mode.
Prole
num.
1-10
Segment
Temp.
ºC
Dell
duration
hh:mm
Ramp
ºC/mi-
nute
Prole
start

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Selecta Incudigit-TFT 2001261 and is the answer not in the manual?

Selecta Incudigit-TFT 2001261 Specifications

General IconGeneral
BrandSelecta
ModelIncudigit-TFT 2001261
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals