SeQual Technologies Inc. Part No: 7040
14
Technical Data
Flow Rate 0.5 to 10.0 LPM
Oxygen Concentration
1.0 to 10.0 LPM 91
3% at sea level
Oxygen Concentration Indicator
Green Light = Normal Operation
Yellow Light Flashing = Below Normal Operation (85%*)
Red Light Flashing = Abnormal Operation (70%**)
Ox
en Outlet Pressure Nominal 6.0 psi (41.4 kPa)
1
- Maximum 11.0 psi (75.8 kPa)
Electrical Powe
115V~, 60Hz, 5.0A Nominal, 220-240V~, 50Hz, 2.2A Nominal
O
eratin
Tem
erature
50° F to 104° F
10°
C to 40° C
Dimensions
H x W x D
26.5 in.
66 cm
H
14.7in.
37 cm
W
19.5 in.
50 cm
D
Audible Alarm Indicators
Power Interruption
Power Surge (simulated circuit breaker)
Compressor Malfunction
Excessive Internal Temperature
Outlet Pressure [35-36 psig (241.2-248.1 kPa) relief valve]
Abnormal Oxygen Flow Rate >±30% target flow
Low Therapeutic Oxygen Output** <70% oxygen
Back-u
Alarm Powe
9V Battery
Filters Cabinet Air Inlet, Compressor Inlet and HEPA
Device Classification
IEC Class II, Type BF Applied Part, Continuous Operation
Transport/Storage Requirements
Temperature: -4° F (-20° C) to +140° F (+60° C)
Humidity: Up to 95% Non-Condensing
Back pressure Effect on
Flow Indicator
0 psig (0 kPa): 10.0 LPM
1 psig (7 kPa): 10.0 LPM Nominal
Temperature: 50° F to 104° F (10° C to 40° C) Independent of
Atmospheric Pressure.
OSCI Operation*
85±3% and 70±5% oxygen concentration measurement accuracy
*Where equipped. Subject to ±3% system concentration accuracy, and for backpressures up to 7kPa.
**Where equipped. Subject to ±5% system concentration accuracy, and for backpressures up to 7kPa.
1
Outlet pressure may decline to 5.0 psig (34.5 kPa) nominal at 10,000 ft altitude.
The system performance may decline above 10,000 ft (3048m).