6 Faults
6 Faults
If a fault occurs during operation, an error message will appear on
the screen. The error number and error name will be shown.
If more than one error occurs, only the error that occurred rst
will be displayed. Each additional error message will be displayed
only after the previous error has been corrected.
Overview of possible error messages
Error
no.
Appearance
on LCD (error
name)
Cause Error correction
9 NO
ADJUSTMENT
No adjustment data
available
Carry out adjustment
10 ADJUSTMENT
ERROR
Zero point in ppm range
(SC)
Check test gas or repeat
adjustment
11 ADJUSTMENT
ERROR
Zero point in LEL range
(CC)
Check test gas or repeat
adjustment
12 ADJUSTMENT
ERROR
Zero point in vol. range
(TC)
Check test gas or repeat
adjustment
13 ADJUSTMENT
ERROR
Sensitivity in ppm range
(SC)
Check test gas or repeat
adjustment
14 ADJUSTMENT
ERROR
Sensitivity in LEL range
(CC)
Check test gas or repeat
adjustment
15 ADJUSTMENT
ERROR
Sensitivity in vol. range
(TC)
Check test gas or repeat
adjustment
40 END OF
RANGE
Measuring range
violation in LEL range
Adjust LEL range
51-54 ERROR
UNKNOWN
Component error Switch device off and back
on or consult SEWERIN-
Service
59 VOLTAGE
SUPPLY
Voltage outside
permissible range
Error can only be corrected
by SEWERIN Service
60 ERROR
UNKNOWN
Semiconductor sensor
break (SC)
Error can only be corrected
by SEWERIN Service
61 ERROR
UNKNOWN
Catalytic combustion
sensor break (CC)
Error can only be corrected
by SEWERIN Service
62 ERROR
UNKNOWN
Thermal conductivity
sensor break (TC)
Error can only be corrected
by SEWERIN Service
100 PUMP
CAPACITY
Insufcient pump
capacity
Check lter in device and
in probes