Overview

The Shenchen Precision Pump LabV Series is an intelligent peristaltic pump designed for flow measurement, suitable for laboratory, equipment support, and industrial production. It features a 4.3-inch color touch screen for control, displaying working state, flow data, setting parameters, and system settings on a single screen. The pump supports intelligent calibration and online micro-adjusting functions, offering three measurement modes: fixed volume measurement, dispensing, and timer start and stop. It is compatible with various pump heads, including EasyPump, AMC (AMC1-AMC12), YZ1515x, YZ2515x easy load, MC series multichannel (MC1-MC12), SN series standard (SN15, SN25), and DZ25-3L.

Function Description:

The LabV series operates as a flow measurement type intelligent peristaltic pump. Its core functions include:

  • Flow Rate Control: Displays current flow rate or motor speed. Users can amend the flow rate or speed, though this is restricted when dispensing is active.
  • Real-time Dynamic Display: Provides a dynamic visual representation of the pump's current running status and results.
  • Real-time Parameter Display: Shows current running state and set parameters. When fixed volume measurement is on, it displays fixed volume parameters; when dispensing is on, it displays dispensing parameters. Otherwise, all parameters are zero.
  • Setting Parameter Display: Shows fixed volume measurement and dispensing state information, including pump head model and tube size.
  • Date and Time Display: Displays current date and time, which can be changed in system settings. An alarm clock icon indicates if the timer start and stop function is active.
  • System Settings: Allows users to configure various parameters.
  • Flow Calibration: Enables calibration of the flow rate for accuracy.
  • Date & Time Setting: Allows setting the current date and time.
  • Fixed Volume Measurement: Activates a mode where the pump automatically measures and transfers a set volume, stopping once the target volume is reached. The flow rate can be adjusted during this process.
  • Dispensing: Allows the pump to transfer a fixed volume within a set time for a specified number of repeat times.
  • Timer Start and Stop: Enables setting specific start and stop times for the pump to operate automatically. This function is unavailable when fixed volume measurement or fixed time and volume functions are active.
  • Back Suction Angle: Prevents liquid from dropping when the motor stops, particularly useful for viscous liquids. The range is 0-360°.
  • External Control: Supports external start/stop, CW/CCW direction control, and speed control via analog signals (0-5V, 0-10V, 4-20mA) or pulse mode. External control modes are independently set and activated.
  • Communication: Supports Modbus-RTU mode via RS485 or RS232 interfaces, allowing communication with a master device. The pump only receives communication control when in the main interface.

Important Technical Specifications:

  • Flow Rate Resolution: 0.01 mL/min
  • Power Supply: AC 220V±10% 50Hz/60Hz (standard); AC 110V±10% 50Hz/60Hz (optional)
  • Operation Mode: Touch screen and mechanical keypad
  • External Control:
    • Switch signal
    • Passive switching signal: Foot pedal switch
    • Active switching signal: 5-24V universal
    • Analog signal input: 0-5V, 0-10V, 4-20mA for option
  • Communication: RS232/RS485, supports Modbus (RTU mode)
  • Output Port: Output motor running status (Open collector output)
  • Back Suction Angle: 0-360°
  • Protection Rate: IP31
  • Speed Range:
    • LabV1, LabV1-II, LabV1-III: 0.1-150 rpm
    • LabV3, LabV3-II, LabV3-III: 0.1-350 rpm
    • LabV6, LabV6-II, LabV6-III: 0.1-600 rpm
  • Power Consumption:
    • LabV1, LabV1-II, LabV1-III, LabV3, LabV3-II, LabV3-III, LabV6: <50W
    • LabV1-III, LabV3-III, LabV6-III: <80W
  • Motor Type:
    • LabV1, LabV1-II, LabV3, LabV3-II, LabV6: Stepper motor
    • LabV1-III, LabV3-III, LabV6-III: Closed loop stepping motor
  • Temperature: 0-40°C
  • Humidity: <80%

Usage Features:

  • User-Friendly Interface: 4.3-inch color touch screen with animation for intuitive operation.
  • High Accuracy: Intelligent calibration and accurate angle control technology ensure high precision for transferring and dispensing liquids.
  • Online Micro Adjusting: Allows real-time adjustment of flow rate during production to compensate for tubing fatigue and maintain accuracy without stopping the process.
  • Versatile Measurement Modes: Supports fixed volume, dispensing, and timed start/stop operations, catering to various laboratory and industrial needs.
  • Pump Head Compatibility: Can load multiple pump heads (single or multichannel) to meet different application requirements due to high torque and low power loss.
  • Memory Function: Power-down memory stores running parameters, ensuring safe and reliable operation.
  • Fast Fluid-Filled Function: Quickly washes tubing and fills it with fluid.
  • External Control Support: Convenient for integration with other equipment, offering start/stop and direction control.
  • Numeric Keypad Input: Provides a clear interface for entering numerical data, with unit switching and clear/backspace functions.

Maintenance Features:

  • Pre-Operation Check: Users should check the machine's running status before starting to ensure normal operation.
  • Leakage and Fault Check: Regular checks for leakage and correction of any appearing faults.
  • Liquid Spill Cleanup: Prompt cleaning of any liquid overflow from the pump.
  • Power-Off for Maintenance: Always turn off the power supply and disconnect the plug (holding the plug, not the wire) if liquid splashes on the pump or if maintenance/repair is needed.
  • Grounding Requirement: The power socket must have a reliable ground wire.
  • External Control Connection Safety: Foot pedal switches and other external control plugs must be connected or disconnected only when the pump is powered off to prevent damage to the external control interface.
  • Water Environment Precaution: The product is not waterproof; protective measures are required when used in wet environments.
  • Special Certifications: The product does not have medical or military certifications; self-certification is required for use in such specialized fields.
  • Long-Term Storage: For long-term storage, the pump should be cleaned and kept in a dry, ventilated environment.
  • Warranty: The pump comes with a 3-year warranty, excluding normal wear and tear, lack of reasonable maintenance, consumable items (tubing, accessories), electrical surges, chemical attacks, improper operation, or man-made damage.

Shenchen Precision Pump LabV Series Specifications

General IconGeneral
BrandShenchen Precision Pump
ModelLabV Series
CategoryWater Pump
LanguageEnglish