4. Installation
Do not open or alter any part of the product, which would lead to malfunction and irrecoverable
damage.
For the installation, the FS8001 has an installation hole fixture that can be fitted with M3 screws.
Proper torque should be applied to avoid damage. The recommended torque for the M3 screw is
0.25±0.03 N m. The sensor comes with a standard data and power cable of 0.5m, users can extend
the cable with a shielded counterpart. Make sure to the mechanical leakage proof of the connections
and all electrical precautions are applied. Please make sure the electrical pins are properly engaged.
It should be noted that the sensor is designed for medium to low pressure per the applications,
therefore, the system design would be important for the flow stability and related flow noises.