SITRANS P300 with HART communication
Operating Instructions, 06/2005, A5E00359579-02
8-1
Commissioning
8
8.1 Safety instructions for commissioning
Warning
Intrinsically safe circuits
With intrinsically-safe circuits, use only certified ammeters appropriate for the transmitter.
Specific type examination certificates
If the transmitter is used in a hazardous area, comply with the EC type examination
certificate or the certification valid in your country.
Warning
"Intrinsically safe"
If a non-conforming supply unit is used, the "intrinsically safe" protection type will no longer
be effective and the approval certification will be invalid.
Warning
Fixed installation
Devices to be operated in hazardous areas should be connected with fixed cable installation.
This is not necessary with intrinsically safe devices or devices with "nL" – "limited energy"
protection.
Warning
Tightness
Use cable with a cable diameter of 7 to 12 mm for the IP65 or IP68 protection rating.

Table of Contents

Other manuals for Siemens 7MF8.24 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens 7MF8.24 Series and is the answer not in the manual?

Siemens 7MF8.24 Series Specifications

General IconGeneral
ManufacturerSiemens
Accuracy±2 mm (0.08 inch)
Output Signal4 to 20 mA HART
Housing MaterialAluminum
Process ConnectionThreaded, Flanged
Electrical ConnectionM20 x 1.5 or 1/2" NPT
Ambient Temperature Range-40°C to +80°C (-40 to +176 °F)