6 - 1
Siemens AG SPR2-310.662.01 Page 1 of 2 ARCADIS Varic
Medical Solutions Rev. 02 05.06 CS PS SP
Technical Data 6
Technical data for DHHS components 6
In accordance with Federal Radiation Performance
Standards 21 CFR Subchapter J Paragraph 1020.30 (h)
X-Ray Controls and Generators 6
Technique factors for maximum line current:
110 kV
p
tube voltage, 13 mA tube current
Generator ratings: 6
6
Nominal
line voltage:
Rated line
voltage:
Line-voltage-
regulation:
Maximum line
current:
100 V ± 10% 11% 28 A
110 V ± 10% 9% 26 A
120 V ± 10% 8% 24 A
127 V ± 10% 7% 22 A
200 V ± 10% 7% 14 A
230 V ± 10% 5% 12 A
240 V ± 10% 5% 12 A
Fluoroscopy: 110 kV
p maximum voltage
15.2 mA maximum current
Pulsed Fluoroscopy: 110 kV
p maximum voltage
23.0 mA maximum current
Digital Radiography: 110 kV
p maximum voltage
23.0 mA maximum current
Digital Radiography (k=1):
(Single Shot Mode)
110 kV
p maximum voltage
250 mA maximum current
Road Map: 110 kV
p maximum voltage
15.2 mA maximum current
Subtraction: 110 kV
p maximum voltage
15.2 mA maximum current
Radiography: Duty cycle: 1:60
- 40 kV/20 mA
- 110 kV/13 mA