Function tests 5 - 9
Siemens AG SPR2-310.662.01 Page 9 of 10 ARCADIS Varic
Medical Solutions Rev. 02 05.06 CS PS SP
7.2 Center cassette to radiation field 5
Work sequence:
1. Use the film pairs already exposed (C-arm in basic position, C-arm to the right and
C-arm to the left) to evaluate the center deviation.
2. Place the film pairs one on top of the other.
3. Mark the geometric center of the 24 cm x 30 cm film.
4. Mark the geometric center of the radiation field.
5. Calculate the deviation.
C-arm basic position:
C-arm to the left, tube assembly at stop
C-arm to the right, tube assembly at stop
If the deviations are within the given tolerance, no additional tests are necessary.
Fig. 8
= .......... mm (must be < 19mm), (2% of the SID)
= .......... mm (must be < 19mm), (2% of the SID)
= .......... mm (must be < 19mm), (2% of the SID)
A
Rad. field
Film 35 cm x 35 cm
Film (cassette 24 cm x 30 cm)

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors

Related product manuals