ARCADIS n.a. SP00-000.841.01 Page 4 of 10 Siemens AG
System Manual Rev. 01 08.05 CS PS 24 Medical Solutions
3 - 4 Installing the ON/OFF assembly
Replacement and wiring of PS/2 cable 3
The PS/2 cable in the monitor trolley is discarded and replaced by the PS/2 cable supplied
with the cable set.
Remove the old PS/2 cable completely.
To do this, cut open the cable ties and remove them.
Run the new PS/2 cable from the PS/2 plug of the PC (imaging system) parallel to the
existing cables of the CIPP board (D66) to the ON/OFF assembly D50 and connect it to
connector D50.X13.
Attach the new cable to the existing ones with cable ties. Make sure there is no
mechanical strain (tension on the cable) on the PS/2 connector.
Wiring of adapter cable D50.X12- X22 3
Wire up the adapter cable D50.X12 - X22 supplied with the cable set.
To to this, connect the plug connection designated X22 with connector X22
and plug the other end D50.X12 into the ON/OFF assembly.
Wiring on D50 3
Plug in connector D50.X9 and screw it on. Make sure that connector D50.X16 is not bent
or disconnected.
Plug in connector D50.X8.
Plug in connector D50.X6.
Plug in connector D50.X7.
Plug in connector D50.X13.

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors

Related product manuals