Batteries
The batteries shall always be sufficiently charged.
Caution
Weak batteries of wireless footswitch
Discontinuation of interventional treatment
Make sure that the batteries are charged before starting an
examination.
In case of empty battery, use the provided power adapter to continue
treatment.
The batteries of the wireless footswitch must be exchanged once a year by
authorized and trained staff. Please call Siemens Service. Unauthorized
exchange may cause severe damage.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user´s authority to operate the equipment.
Cleaning and disinfection
Caution
Use of harsh cleaning agents, liquids or sprays.
Risk of electrical hazard or damage to the system
Use only substances for cleaning and disinfection, which are
recommended.
Do not let cleaning liquids seep into the openings of the system, e.g. air
openings, gaps between covers.
Observe the following cleaning and disinfection instructions.
Cleaning agents for all LCD displays, Large Display and Artis
Cockpit only
The following active ingredient classes can be used for LCD
displays:
Active ingredient class Tested cleaning agents and disinfec-
tants
Additional examples
Alcohol Ethyl alcohol 96% Hospiset cloth
Microcide liquid
Aldehyde Melsitt
Cidex
Aldasan 2000
Kohsolin
Gigasept FF
Chlorine derivates Terralin Quartamon Med
2.4
2.4.1
2 Safety
40
Operator Manual
Print No. AXA4-100.620.17.01.02