6 Rotate the coil by 180 degrees so that the center marking of the
coil (embossed cross) points upward.
7 Connect the coil plug to a suitable socket.
8 Align the laser-light localizer with the center marking of the coil.
9 Start the quality measurement.
5 Quality measurement
20 Contour 24 Coil | Operator Manual
Print No. MR-06032G.640.01.02.02