D
donna05Aug 11, 2025
Why is there a line parallel to the short axis on my Siemens MAMMOMAT Novation DR?
- MmosleysaraAug 11, 2025
If you observe a line parallel to the short axis on your Siemens Medical Equipment, it could be due to several reasons related to seam correction and detector calibration. Specifically, the seam correction algorithm might not be turned on or the wrong algorithm might be selected. Another cause could be that the detector gain isn't calibrated properly, or the seam (lines 1535 and 1536) needs mapping. If the issue stems from source lines, it might be due to improper detector calibration or the line needing mapping. To address this: * Perform the detector calibration procedure. * Perform the bad pixel correction.