Siemens AG SPB7-250.815.06.02.02 MAMMOMAT Novation DR
12.05 CS SD 24
Mechanical covers 33
Page 33 of 116
Medical Solutions
Fig. 15: Covers and sign
1. Mount the side covers (1/Fig. 15 / p. 33) with six screws (2/Fig. 15 / p. 33) on each side.
Start with the right side.
NOTE
Do not forget to fit the sign when mounting the side covers (see
(3/Fig. 15 / p. 33)).
Use the two longer screws.
NOTE
Make sure that the longer screws are not used at the back of the
covers at the upper location since this may damage the board un-
derneath the cover (see (6/Fig. 15 / p. 33)).
Front cover 0
CAUTION
Risk of damages.
If the covers are exposed to internal stress, cracks might arise.
¹ The following work must be carried out with caution.
1. The side covers must be flush at the front. If necessary, loosen the screws and adjust
the side covers.
2. Carefully fit the front cover (see (4/Fig. 15 / p. 33) so that both openings engage with
the lugs of the side covers.
3. Carefully swing the front cover upwards and let it snap in position over the side covers.