Siemens AG SPB7-250.815.06.02.02 MAMMOMAT Novation DR
12.05 CS SD 24
Important Notes 15
Page 15 of 116
Medical Solutions
Delay times between two exposures 0
The delay times listed below must be followed to prevent the tube from overheating.
Tab. 4 Delay times between two exposures
Exposure mAs value Delay time between two exposures (seconds)
max. 100 min. 15
max. 200 min. 30
max. 300 min. 45
max. 400 min. 60
max. 500 min. 75