104 Start-up Protocol
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 104 of 116
Medical Solutions
Tab. 21 AEC Dose settings
OPDOSE settings 0
Results from the chapter "Checking OPDOSE" on (Checking OPDOSE / p. 57).
Tab. 22 OPDOSE Settings
Check of grid lines 0
Results from the chapter "Checking Grid Lines" on (Checking Grid Lines / p. 63).
Tab. 23 Measuring Grid Lines
AEC kV Anode /
Filter
Dose value Pixel value
H
28 kV
Mo/Mo
D
H
Mo/Rh
-
D-
H
W/Rh
-
D-
Program
button
Thickness kV
Anode/filter
combination
AEC
1
2
3
4
Exposure Exp Mode mAs PMMA/
mm
Tube
Angle
visible
grid
lines
(yes/no)
X Y Mean pixel
value differ-
ence [%]
1 mAs 32 30 0
o
2 mAs 56 40 0
o
3mAs200600
o
4AEC "H"-400
o
5 AEC "H" - 40 + 60
o
6AEC "H"-40- 60
o