58 Checking dose settings
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 58 of 116
Medical Solutions
Checking the AEC dose settings 0
The dose values for the AEC settings (H & D selection on the control console) have to be
checked.
As a default from the factory, the following BRICK / AEC Configuration values are set
(for further details see the Software Brick V... manual SPB7-250.816.04...):
H = (80, 70, 45)
D = (100, 80, 60)
Checking the H & D dose settings for Mo/Mo.
Proceed as follows:
1. Mount the collimator mounted plexi phantom and the compression plate simulator.
2. Register or create a patient and submit with Exam.
Fig. 38: Calibration_Patient register