96 Final procedures
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 96 of 116
Medical Solutions
Protocols 0
• The manufacturer of this product requires information and the legislator demands
proofs that a product delivered free of defects from the factory continues to possess the
required and certified quality properties on installation and start-up.It is therefore abso-
lutely necessary that the installation Protocol with the installation and start-up data is
sent without delay after completion of the work to the address stated on the installation
protocol.
• The SP Q ’Installation Protocol’ has to be filled in and signed. The form can be found in
chapter 9 ’Certificates’ of the technical documentation binder (Appendix of 66 46 900
QZE 01S xx). It has to be sent to the address stated on the protocol.
• You will find the protocol in the system folder Place a copy of the protocols in the techni-
cal documentation binder.
• The Startup Protocol and the Installation Protocol remain with the system.