92 Final procedures
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 92 of 116
Medical Solutions
Service PC and measuring instruments 0
Saving the configuration file 0
NOTE
Make sure that the backup floppy is inserted.
Main menu:
1. Select: Configuration -> Save config file.
2. Press F2 to save the data.
Reset the exposure counter
Main menu:
1. Select: Service -> Reset exposure counter.
2. Delete the error memory with Y.
Deleting the error buffer
Main menu:
1. Select: Service -> Delete error buffer.
2. Delete the error memory with Y.
Saving the MAMMOMAT Parameters
Main menu
1. Select Backup -> Copy Installation Area to Floppy -> Stand.
2. Select Backup -> Copy Installation Area to Floppy -> Panel.
3. Exit the program with F10 when the backup is completed.
NOTE
Keep the backup floppy with other documentation for the MAM-
MOMAT Novation
DR
in question.