42 Mechanical covers
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 42 of 116
Medical Solutions
7. Set the compression to 7kg by using the presetting knob (5/Fig. 23 / p. 41).
The preset compression force is shown on the display at the lower part of the stand
(1/Fig. 25 / p. 42).
Fig. 25: Display at the lower part of the stand_2
8. Press the foot switch for motorized compression to move the compression plate down-
wards while simultaneously pressing the compression plate in the opposite direction by
hand.
The compression plate should now slow down and stop when the preset compression
force is reached.
NOTE
During compression, the motorized vertical adjustment and the ro-
tary motion of the swivel arm system should be blocked.
Decompression button 0
Checking the OPCOMP 0
1. Decompress.
2. Set the compression force to 20 kg by using the presetting knob (5/Fig. 23 / p. 41) and
compress.
The compression movement must be interrupted by OPCOMP before 20 kg is reached,
i.e. at 5-15 kg depending on the material (4/Fig. 23 / p. 41) used.
Check the decompression button on the separate control console for proper
function, i.e. that the compression plate moves upwards when the button is pressed.