Siemens AG SPB7-250.815.06.02.02 MAMMOMAT Novation DR
12.05 CS SD 24
Detector installation 27
Page 27 of 116
Medical Solutions
Mounting the detector 0
Fig. 11: Detector installation
1. Remove the micro switches and their holders S2 and S3.
2. Remove the cover from the detector cable connector.
3. Loosen the grid holder, 2 screws (4/Fig. 11 / p. 27).
4. Attach the detector with the four bolts (1/Fig. 11 / p. 27). Make sure that the bolts are
tightened.
NOTE
To be sure that the detector is correctly mounted and that it is lo-
cated directly below the radiation field, use the guiding lines on the
wing and make sure that the detector is in the right position before
screwing the bolts.
5. Insert the detector cable underneath the grid holder and tighten the grid holder screws
again (4/Fig. 11 / p. 27).
6. Connect the detector cable (3/Fig. 11 / p. 27) to the cable coming out of the wing
(2/Fig. 11 / p. 27).
7. Turn the actuating screw until it is finger-tight only. The remaining gap should not ex-
ceed 2mm.
8. Check that the pins on the connectors have not been pushed back.
9. Mount the cover over the detector cable with 2 screws. Make sure that no cables are
squeezed.