40 Mechanical covers
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 40 of 116
Medical Solutions
Checking the field light time 0
Fig. 21: Field light switch
1. Switch on the field light by pressing the switch (1/Fig. 21 / p. 40) or by pressing the
compression foot switch (1/Fig. 22 / p. 40).
The field light should go out automatically after a preset period of time (normally 20 s.)
Fig. 22: Footswitch
Adjusting the field light time
The illumination time of the light field can be adjusted within a certain range according to
the customer“s requirements.
NOTE
A shorter illumination time results in an increased lifetime of the
lamp.
1. Connect the service PC to the D750 and start the MAMMOMAT Service Software.
2. In the Main menu select Configuration -> Miscellaneous -> Illumination time.
3. Type in the illumination time and save with <F2>.