Siemens AG SPB7-250.815.06.02.02 MAMMOMAT Novation DR
12.05 CS SD 24
Mechanical covers 39
Page 39 of 116
Medical Solutions
Vertical adjustment
CAUTION
Remove the protective strips before performing the vertical adjust-
ment of the swivel arm.
If not removed, the protective strips could be damaged.
¹ Make sure the protective strips are removed before
adjustments, see "Metal Curtain" on (Metal
curtain / p. 12)
1. Press the buttons for upward and downward movement ((3/Fig. 19 / p. 37)) of the swiv-
el arm system.
The movement should stop automatically when either the end position is reached or
when the button is released.
NOTE
All eight vertical adjustment buttons (four on either side
(3/Fig. 19 / p. 37)) must be checked.
The switching off in upper and lower end position is effected by
switches S881 and S884 respectively.
Emergency stop
1. Press the emergency stop button (4/Fig. 19 / p. 37).
The button should latch and all motorized movements of the equipment should be
blocked ("Er 813" should appear). To reset the emergency stop button, turn it
clockwise.