38 Mechanical covers
MAMMOMAT Novation DR SPB7-250.815.06.02.02 Siemens AG
12.05 CS SD 24
Page 38 of 116
Medical Solutions
5. Check the rotary motion of the swivel arm system by pressing the buttons
(2/Fig. 19 / p. 37) one by one.
The rotation of the swivel arm system should stop when the button is released or when
the end position is reached (+135 degrees clockwise and -180 degrees
counterclockwise). The projection angle is shown on the display at the lower part of
the stand.
Fig. 20: Display at the lower part of the stand
6. Set the projection angle to 60 degrees with the presettings knob (1/Fig. 19 / p. 37).
The preset angle is displayed for approximately 3 s.
7. Press one of the buttons (2/Fig. 19 / p. 37).
The rotary motion of the swivel arm system should stop automatically when the preset
projection angle (both + and -) is reached and at 0 degrees, +90 degrees or -90
degrees.
NOTE
All eight rotary motion buttons (four on either side) must be
checked.(2/Fig. 19 / p. 37)).
After the button has been released, the swivel arm system should
remain in the position set.
OPDIMA wing
1. Select the OPDIMA wing; OP is then displayed on the control console.
2. Set the projection angle to minimum with presetting knob (1/Fig. 19 / p. 37).
3. Check the rotary motion of the swivel arm system by pressing the buttons
(2/Fig. 19 / p. 37) one by one.
The rotation of the swivel arm system should stop when the button is released or when
the end position is reached (+120 degrees clockwise and -120 degrees
counterclockwise). The projection angle is shown on the display at the lower part of
the stand.