Siemens AG SPB7-250.815.06.02.02 MAMMOMAT Novation DR
12.05 CS SD 24
Power up the MAMMOMAT 19
Page 19 of 116
Medical Solutions
Mains voltage 0
WARNING
Before the power for the MAMMOMAT is turned on, make sure that
the cable plugs (left and right of the system) for the magnification
table are not shorted.
¹
The mains input can be either 1-phase or 2-phase, 50 or 60 Hz.
Electrical power to mains input: 0
Mains voltage 1-phase:
or
2-phase:
208, 230, 240 and 277 V ±10%
208, 230, 240, 277 and 400 V ±10%
Mains Frequency: 50 or 60 Hz
Maximum load: 10kVA (momentary) and 0.8 kVA (long time load).
Fuse: 25 A (external) 208 -> 400 V.