MOBILETT Plus/ Plus E Register 6 SPR8-215.850.02 Page 2 of 2 Siemens-Elema AB
Rev. 01 01.02 SPS-UD Solna, Sweden
6 - 2 Power-up sequences
The oil temperature in the single tank is measured. If the temperature is not within the
range +8°C to +50°C, an error will be displayed.
The positions of service switches S1 and S2 on the CPU board D1 are queried.
The switch positions are:
S1 and S2 off = normal mode
S1 on = software-supported service mode
S2 on = PC mode
PC mode is used only for testing procedures at the factory!
The exposure parameter display indicates PC-Mode and the selection buttons on the
Control and display panel are disabled!
The charging procedure for the capacitor bank is completed after reaching the final
voltage value. According to the residual voltage in the capacitor bank, this takes between
5 and 120 s.
The green ready LED on the Control and display panel is turned on.
The unit is ready to operate!