MOBILETT Plus/ Plus E Register 6 SPR8-215.850.02 Page 2 of 2 Siemens-Elema AB
Rev. 01 01.02 SPS-UD Solna, Sweden
12 - 2 Preparation and exposure release
Exposure release 12
The secondary contact in the exposure-release switch is closed. This initiates the
following additional functions:
• the yellow exposure-indicator lamp on the Control and display panel is turned on,
• the high-voltage inverter is switched on,
• the resulting tube current is integrated until the preset mAs value is reached, followed by;
• the high-voltage inverter is switched off,
• preheating is resumed
• a long beep tone is generated.
• the exposure-indicator lamp is turned off.
• the rotating anode is braked.
If the MOBILETT Plus is equiped withe a remote
exposure switch system, please see Installation
Instruction: Installation of remote exposure switch
system, chapter 3 "Supplement to the Instructions for
Use".