15 - 1
Siemens-Elema AB Register 6 SPR8-215.850.02 Page 1 of 4 MOBILETT Plus/ Plus E
Solna, Sweden Rev. 01 01.02 SPS-UD
High-voltage circuit 15
The block diagrams and the wiring diagram X037E, sheet 140 are the basis for the func-
tional description given below.
The high voltage circuit includes:
the capacitor bank, M1,
the CPU board D1,
the high-voltage inverter board D13,
the single tank H1
The high-voltage inverter on board D13 includes eight power MOS-FETs and functions as
a square wave inverter.
It is supplied with power from the capacitor bank, M1, with power and with control pulses
in the frequency range 20 - 50 kHz to control its function from microprocessor SAB
80C166 on the CPU board D1.
The pulse-pause ratio of this control frequency is modulated by the microprocessor on the
CPU board D1.
It is possible to measure the control frequency at the test points INV A and INV B on
board D13.
The high-voltage inverter drives a DC voltage of alternating polarity via the primary wind-
ing of the single tank H1.
The secondary winding of the single tank H1 transforms these current pulses into high
voltage and a voltage doubler circuit generates their final value.
Fig. 1