MOBILETT Plus/ Plus E Register 6 SPR8-215.850.02 Page 2 of 2 Siemens-Elema AB
Rev. 01 01.02 SPS-UD Solna, Sweden
18 - 2 Error messages
All of these messages are stored in a so - called Error-history memory, where
they are available at any time for later analyses during servicing.
In the error-history memory a maximum of 20 error messages can be stored.
The error history memory can be called up with the service program P08 and cleared
with program P09.
The selection procedure and details concerning the service programs mentioned are
given in the Service Instructions.