EasyManua.ls Logo

Siemens MOBILETT XP Hybrid - Page 65

Siemens MOBILETT XP Hybrid
109 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Radiation Calculation
MOBILETT
SPR8-230.621.01.03.02 63
of 106
Calculate the dose area product with Equation (1):
DAP = 57 (µGy x m
2
/mAs) x 10 x 10 x 10 (mAs) x 6.40 (cm
2
/m
2
)
= 364800 µGycm
2
Since 1000 µGy = 1 mGy, DAP = 364 800 µGycm
2
365 mGycm
2

Table of Contents

Other manuals for Siemens MOBILETT XP Hybrid

Related product manuals