Pre-installation Instructions
Multimobil 5C
Siemens Ltd. Med India Version 5.0 Page 4 of 11
Copyright © SIEMENS LTD. All rights reserved. For internal use only
Sr.No Parameter Value
reset and 5 minute buzzer indication
18. Thermal Protection Auto cutoff of fluoroscopy and
radiography in case of tube head
temperature exceeding 60°(+5°)
Celsius
19.
Automatic termination
Buzzer alarm is set and fluoroscopy
terminates after 10 minutes of
Continuous uninterrupted fluoroscopy.
20. Power Input:
Continous
Instantaneous
1.5kVA
3.5kVA
21. Mains Isolation
External: done by the
user
Power supply cord shall be plugged
where both poles (L & N) are isolated
simultaneously.
22. Type and degree of
protection against electric
shock (EN 60601 – 1)
Class – I, Type B equipment
23. X - ray Tube Stationary Anode DF-151R
24. Focal Spot – nominal
value
0.5 for Fluoroscopy
1.5 for radiography
25. Inherent tube Filtration 0.8 mm Al at 50kV
26. Nominal kV 110 kV
27. Target Angle
16°
28. Collimator Motorised, multileaf Iris Collimator
29. Total filtration 5.3 mm Al at 100kV
30. Exposure release Switch.
Flouroscopy switch.
Hand held - 1 Step.
Foot switch – 1 step and / or
Hand held switch – 1 step.
31. Exposure Rate for
Radiography
Pulse to pause ratio 1:30; corresponds
to a cool down period of 4 minutes at
maximum output.
32. Max. X-ray coverage 430X430mm at SID 909 mm
33. Mode of Operation Continuous operation
34. Focus-II distance 860mm for 9’’ system
900mm for 6’’ system
35. Max. cassette size at 900
cm SID
10’’X12’’