EasyManua.ls Logo

Siemens MULTIMOBIL 2.5 Service Manual

Siemens MULTIMOBIL 2.5
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
MULTIMOBIL 2.5
English
Print No.: SPR8-X01.831.01.02.02 Doc. Gen. Date: 09.05
Replaces: SPR8-X01.831.01.01.02
© Siemens AG 2003
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
SP
Maintenance Instructions

Other manuals for Siemens MULTIMOBIL 2.5

Question and Answer IconNeed help?

Do you have a question about the Siemens MULTIMOBIL 2.5 and is the answer not in the manual?

Siemens MULTIMOBIL 2.5 Specifications

General IconGeneral
BrandSiemens
ModelMULTIMOBIL 2.5
CategoryMedical Equipment
LanguageEnglish

Summary

Maintenance Instructions

General

General safety and performance checks required every 12 months for proper operation and safety.

Replacing Damaged or Missing Screws

Instructions for replacing screws, emphasizing ESD regulations and specified screw tensile strength.

Cleaning

Guidelines for cleaning the unit safely, avoiding specific solvents, organic solvents, or spray methods.

Product Disposal

Legal regulations for product disposal and informing Siemens Customer Service to avoid environmental damage.

Electrolytic Capacitors Disposal

Instructions for emptying electrolytic capacitors and contacting customer service for proper disposal.

Mechanical Risk

Addresses mechanical risks, specifically the tension in the spring unit, and advises contacting customer service.

Safety Checks

User must follow daily and monthly safety checks to ensure proper functioning and safety.

Daily Safety Checks

Lists daily checks: power cord condition, exposure indicator, locking mechanism, and arm system position.

Monthly Safety Checks

Lists monthly checks: arm movements, locking lever function, and label legibility.

Performance Check (Every Six Months)

Six-month performance checks: counterbalancing, locking, brake, and sample exposure.

Aligning Light Field to Radiation Field

Procedure to align the light field to the radiation field, including operating sequence and adjustment evaluation.

Checking and Readjusting Counterweight

How to check and adjust the counterweight for smooth arm movements and proper positioning.

Adjusting Friction in the Arm

Instructions on adjusting the clamping knob for arm friction and tightening a related nut.

Removing the Top Panel

Steps to remove the top panel, including disconnecting power and securing the unit.

Removing the Front Cover

Steps to remove the front cover, including securing the unit and using soft bedding.

Removing the X-ray Source Assembly

Steps to remove the X-ray source assembly, including disconnecting power and securing movement.

Replacing the Single Tank

Procedure to remove and install a new single tank, including safety checks and connections.

Replacing the Collimator

Steps to replace the collimator, including disconnecting cables and removing screws.

Replacing the Collimator Lamp

Instructions for replacing the collimator lamp, including safety precautions and checking function.

Related product manuals