Do you have a question about the Siemens MULTIMOBIL 2.5 and is the answer not in the manual?
Brand | Siemens |
---|---|
Model | MULTIMOBIL 2.5 |
Category | Medical Equipment |
Language | English |
General safety and performance checks required every 12 months for proper operation and safety.
Instructions for replacing screws, emphasizing ESD regulations and specified screw tensile strength.
Guidelines for cleaning the unit safely, avoiding specific solvents, organic solvents, or spray methods.
Legal regulations for product disposal and informing Siemens Customer Service to avoid environmental damage.
Instructions for emptying electrolytic capacitors and contacting customer service for proper disposal.
Addresses mechanical risks, specifically the tension in the spring unit, and advises contacting customer service.
User must follow daily and monthly safety checks to ensure proper functioning and safety.
Lists daily checks: power cord condition, exposure indicator, locking mechanism, and arm system position.
Lists monthly checks: arm movements, locking lever function, and label legibility.
Six-month performance checks: counterbalancing, locking, brake, and sample exposure.
Procedure to align the light field to the radiation field, including operating sequence and adjustment evaluation.
How to check and adjust the counterweight for smooth arm movements and proper positioning.
Instructions on adjusting the clamping knob for arm friction and tightening a related nut.
Steps to remove the top panel, including disconnecting power and securing the unit.
Steps to remove the front cover, including securing the unit and using soft bedding.
Steps to remove the X-ray source assembly, including disconnecting power and securing movement.
Procedure to remove and install a new single tank, including safety checks and connections.
Steps to replace the collimator, including disconnecting cables and removing screws.
Instructions for replacing the collimator lamp, including safety precautions and checking function.