Supplement 1 for Operating Instructions C79000-G52xx-C216-03
16 A5E02769382-03, 03/2012
Sequence parameters
The zero gas purge time is twice the AUTOCAL purge time. The pre-purging phase is equal to the AUTOCAL purge time.
The AUTOCAL purge time is an adjustable parameter; its setting is described in the operating instructions, C79000-G52xx-
C216, section 5.8.4.
The following illustration show how the protection function runs over time:
Concentration
Function test
Zero gas valve open
Zero gas valve closed
ASZ = Autocal purge time (for H
2
S usually: 5 min.)
MW = measured value
NG = zero gas
12.5 vpm
MW > Limit
2 * ASZ
Pre-purging
1 * ASZ
NG purging
2 * ASZ
MW > Limit
2 * ASZ
t
Figure 5 The H
2
S purging function over time