Uroskop ACCESS SPL5-330.840.01 Page 10 of 12 Siemens AG
Endoscopy Option Rev. 01 03.03 CS SD 24 Medical Solutions
5 - 10 Error Correction
Example of the test configuration with a chrome-plated Allen wrench, contrast OK.
Arrow 1:
The reflective areas of the Allen wrench appear almost bright and glaring and have a
noticeable white border, indicating the beginning of camera overmodulation. The reflective
areas still indicate noise in the image.
Arrow 2:
When the monitor is set correctly, the base structures are visible without saturation
effects.
Fig. 8