Page EN-4 XCT-8 Sanitary Transducer– INSTRUCTION MANUAL 7ML19985HX62
English
Outline and Dimensions
Ferrule Considerations
The primary customer requirement is
a 4" sanitary ferrule welded to a short
standpipe on the top of the tank. The
inner standpipe wall and ends must
be smooth and free of burrs, ridges,
or seams, and
must comply with all
applicable regulations.
The standpipe extension between
the mating ferrule and the tank top
cannot be longer than the blanking
distance as determined by the
Siemens Milltronics transceiver. The
blanking distance is the minimum
possible measurement range listed in
the
Specifications
on page 3.
Do not over tighten the mounting.
Hand tightening of the mounting
hardware is sufficient.
1” BSP
or
1” NPT
122 mm
(4.8”)
119 mm
(4.68”)
radiating PTFE face
XCT-8
transducer
sanitary
clamp
standpipe
topped with 4
ferrule
blanking
distance