EasyManuals Logo

Sievers TOC 820 User Manual

Default Icon
255 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #100 background imageLoading...
Page #100 background image
Ionics Instruments 2004 Page 6-13 DLM 30007-08 Rev. A
With the INSTALL option highlighted, press ENTER to view the acid status:
ACID INSTALLATION
ENTER
CLEAR
100% REMAINING
INSTALL DT. : 07/04/92
CURRENT DT. : 07/08/92
ENT = SET W/CURRENT DT.
CLR = EXIT
FIGURE 6-16: ACID Status and INSTALLATION Display
The amount of acid consumed is monitored by the microprocessor and the
percentage remaining is displayed along with the date that the acid reservoir
was installed. Press CLEAR to exit the menu without changing the install date.
For each consumable, there is a means for resetting the percentage left; for
the oxidizer, the installation date may also be reset. These functions are used
in the unlikely event that these values are lost from the battery-backed RAM or
when a new EEPROM chip is installed. For example, the SET % LEFT option
under the ACID INSTALLATION menu can be used to enter the amount of
acid left so that the analyzer will know the volume of acid remaining in the
reagent container. These options are only used for emergencies and the
INSTALL option is used for setting the install dates when the consumables are
replaced.
From the CONSUMABLES menu, print a report of the consumables status by
pressing ENTER with the Print Report option highlighted. The PRINTING
CONSUMABLES screen is displayed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sievers TOC 820 and is the answer not in the manual?

Sievers TOC 820 Specifications

General IconGeneral
BrandSievers
ModelTOC 820
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals