Frequency Counter Characteristics
rms
MHz
100mV
rms
5 Vpp
AC coupling, < 100 MHz
200mV
rms
5 Vpp
AC coupling, 100 MHz ~ 200
MHz
Input
impedance
1M
Ω
Reference Clock Input/Output
Reference Clock Input
Parameter
Min.
Typ.
Max.
Unit
Condition
Frequency
10M
Hz
Amplitude
1.4
Vpp
Input
impedance
5
AC coupling
Reference Clock Output
Parameter
Min.
Typ.
Max.
Unit
Condition
Frequency
10M
Hz
Synchronized to internal
reference clock
Amplitude
2
3.3
Vpp
HiZ load
Output
impedance
50
Ω
Auxiliary In/Out Characteristics
Trigger Input
Parameter
Min.
Typ.
Max.
Unit
Condition
V
IH
2
5.5
V
V
IL
-0.5
0.8
V
Input
impedance
100
Pulse width
100
ns
Response
time
100
ns
Sweep
600
ns
Burst
Trigger Output
Parameter
Min.
Typ.
Max.
Unit
Condition
V
OH
3.8
V
I
OH
= 8 mA
V
OL
0.44
V
I
OL
= 8 mA
Output
impedance
100
Ω
Frequency
1
MHz
Sync Output
Parameter
Min.
Typ.
Max.
Unit
Condition

Other manuals for SIGLENT UM0202X-E02C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SIGLENT UM0202X-E02C and is the answer not in the manual?

SIGLENT UM0202X-E02C Specifications

General IconGeneral
BrandSIGLENT
ModelUM0202X-E02C
CategoryInverter
LanguageEnglish

Related product manuals