99
C4 Manual Rev 7.5.2
GT - Go To Temperature
OUTPUT 701 "GT 85.0" or
OUTPUT 701 "GT 85" or
OUTPUT 701 "GT85"
DL - Delay
OUTPUT 701 "DL 00,25"
OUTPUT 701 "DL00,25"
RA - Ramp to Temperature
OUTPUT 701 "RA 85.5,00,25"
PNn - Select Active Probe
OUTPUT 701 "PN 2"
PTn - Read Temperature
OUTPUT 701 "PT1"
To receive temperature, use the following:
WAIT 200 (this is required)
ENTER 701;A$
DISP A$
The C4 responds only to the following standard I-O statements:
ABORTIO 7 (Aborts all bus activity and resets the C4)
CLEAR 7 (Aborts all bus activity and resets the C4)
CLEAR 701 (Selected device clear, resets the C4)
SPOLL (701) (Polls device #1 for status register contents)
The SRQ must be enabled, using "ES", and the SRQ line must
be activated before the C4 will respond to a serial poll request.