90
C4 Manual Rev 7.5.2
Step
Number
Displayed
Step .
Substep
Keys to press Description of entry
0
00.0
CLEAR/ENTRY 45.2 ENTER
Go to 45.2/C
00.1
CLEAR/ENTRY 5 ENTER
Use 5 minute ramp time
00.2
CLEAR/ENTRY 2 ENTER
Hold temperature for 2 minutes
00.3
CLEAR/ENTRY 1 ENTER
Execute step 1 next
00.4
CLEAR/ENTRY 1 ENTER
Use probe 1
1
01.0
CLEAR/ENTRY 32.3 ENTER
Go to 32.3/C
01.1
CLEAR/ENTRY 2 ENTER
Use 2 minute ramp time
01.2
CLEAR/ENTRY 6 ENTER
Hold 32.3/C for 6 minutes
01.3
CLEAR/ENTRY 2 ENTER
Execute step 2 next
01.4
CLEAR/ENTRY 1 ENTER
Use probe 1
2
02.0
CLEAR/ENTRY 50.1 ENTER
Go to 50.1/C
02.1
CLEAR/ENTRY 0 ENTER
Ramp as quickly as possible
02.2
CLEAR/ENTRY 9900 ENTER
Do not hold at this temperature
02.3
CLEAR/ENTRY 3 ENTER
Execute step 3 next
02.4
CLEAR/ENTRY 1 ENTER
Use probe 1
3
03.0
CLEAR/ENTRY -10.0
Go to -10/C
03.1
CLEAR/ENTRY 0 ENTER
Ramp as quickly as possible
03.2
CLEAR/ENTRY 3 ENTER
Hold -10/C for 3 minutes
03.3
CLEAR/ENTRY 100 ENTER
End program
03.4
CLEAR/ENTRY 1 ENTER
Use probe 1