Using the Centrifuge
Version 05/2010, Rev. 1.8 of 01/09/2011- sb 25
Translation of the original operating manual
6.2.2.3 Installation of Accessories
Only use inserts that are suitable for the rotor (see 11.1 "Suitable
Accessories").
Always load the opposite inserts/buckets of the rotors with the
same accessories and fill to avoid imbalance.
Centrifugation with low capacity: The tubes must be installed
symmetrically so that the buckets and their inserts are loaded evenly.
fig. 6.2: Permissible loading of an angle rotor
6.2.2.4 Tubes
Load the tubes outside of the centrifuge. Liquids in the bores of
the rotor cause corrosion .
Fill the tubes carefully and arrange them according to their
weight. Imbalances result in the excessive wear of the bearings.
Observe the instructions on safety and hazards in chapter 3!