6.2.2.3 Installation of Accessories
• Only use inserts that are suitable for the rotor (see 11.1 "Suitable
Accessories").
• Always load the opposite inserts/buckets of the rotors with the
same accessories and fill to avoid imbalance.
Centrifugation with low capacity: The tubes must be installed
symmetrically so that the buckets and their inserts are loaded evenly.
fig. 6.2: Permissible loading of an angle rotor
6.2.2.4 Tubes
• Load the tubes outside of the centrifuge. Liquids in the bores of
the rotor cause corrosion .
• Fill the tubes carefully and arrange them according to their
weight. Imbalances result in the excessive wear of the bearings.
• Observe the instructions on safety and hazards in chapter 3!