1-14K
Malfunctions and Error Correction
34 Version 05/2010, Rev. 1.8 of 01/09/2011- sb
Translation of the original operating manual
7.2 Error Codes
Error no. Kind of error Measure Note
1-9 System error
• allow to slow down
• power off/on
All these errors stop the
centrifuge or cause it to
decelerate brakeless
10-19 Speedometer error
• allow to slow down
• power off/on
20-29 Motor error
• power off/on
• ensure ventilation
30-39 EEPROM error
• allow to slow down
• power off/on
With error 34, 35, 36, the
centrifuge will stop;
with error 37, 38 error
message only
40-45 Temperature error
(only for refrigerated
centrifuges)
• allow to slow down
• power off/on
• allow to cool down
• provide better ventilation
46-49 Imbalance error
• allow to slow down
• power off
• eliminate the imbalance
50-59 Lid error
• press lid key
• close lid
• remove foreign matter from the
opening of the lid lock device
With error 50 and 51, the
centrifuge will stop
60-69 Process error
• allow to slow down
• power off/on
With error 60 message
"power failure during run",
with error 61, message "stop
after power on"
70-79 Communication error
• allow to slow down
• power off/on
80-89 Parameter error
• allow to slow down
• power off/on
• provide for better ventilation
With error 83, error
message only
90-99 Other errors
• check connections
Fig. 7.4: Error codes from software version 009
If it is impossible to eliminate the errors, contact the service!