DATALOGGER Management
The DataLogger can store up to 60,000 measurement points
(30.000 if also the temperature is stored) in steps ranging from
1s to 10 hours according to the parameter t1 defined in the
Setting Menu. Cycle length is determined by the parameter t2
always defined in the Setting Menu.
Data stored during the last Datalogging are permanently saved in
nonvolatile memory within the gauge so that the measures will
always be accessible until the creation of a new cycle of
measurement
Note:
The functionality of the DataLogger is fully manageable from the
PC using the AnalyzerLight software through which you can:
• View the status of the current cycle.
• Download all the measurements.
• Save all measurements on a file.
• Display the test curve.
• Print the test curve.
• Export to Microsoft Excel the test curve.
• Make the START / STOP of a cycle.
• Set the parameters t1 and t2 and temperature
ON/OFF
.
CYCLE START:
Press simultaneously and hold for a few
seconds the keys e .
The acceptance of START will be shown
on the display by the presence of the icon
SP1.
Each time a measurement point is saved,
the icon SP2 will flash for a one second.
CYCLE STOP:
The cycle will stop automatically at the set
time t2.
Alternatively hold down simultaneously,
for a few seconds, the keys e .
The icon SP1 will be turned off.
VIEW
DATA:
Press and hold simultaneously for a few
seconds, the SET and keys.
The acceptance of the setting will be
displayed on the display by flashing SP1
and SP2.
It is possible now to see all points of
measurement using the button .
To go back of a measurement point use
the key .
To exit from the View Cycle page press
the SET button