Sirona ORTHOPHOS XG 5 Service Manual

Sirona ORTHOPHOS XG 5
630 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
båÖäáëÜ
loqelmelp=ud=Pa
êÉ~Çó
=L=`ÉéÜ
loqelmelp=ud
mäìë
=ap=L=`ÉéÜ
loqelmelp=ud=R=EapF=L=`ÉéÜ
loqelmelp=ud=P=EapF
pÉêîáÅÉ=j~åì~ä
sÉêëáçå=OSKM
ORTHOPHOS XG 3D
ready
/ XG
Plus
Easypad
ORTHOPHOS XG 5 / 3 / 3 PPE
T
R
Prog.
S
kV
mA
P1
Multipad

Table of Contents

Other manuals for Sirona ORTHOPHOS XG 5

Question and Answer IconNeed help?

Do you have a question about the Sirona ORTHOPHOS XG 5 and is the answer not in the manual?

Sirona ORTHOPHOS XG 5 Specifications

General IconGeneral
BrandSirona
ModelORTHOPHOS XG 5
CategoryMedical Equipment
LanguageEnglish

Summary

General Information

Safety

Crucial safety guidelines for handling the device and potential hazards.

Software update

Important information and procedures for performing software updates on the unit.

Messages

Help messages

Explains help codes (Hxxx) caused by operator errors and required actions to resolve them.

Error messages

Details error codes (Ex yy zz) displayed on the Easypad/Multipad, providing error type, location, and troubleshooting info.

List of error messages

Comprehensive list of error codes, their descriptions, required actions, and corresponding page references.

Troubleshooting

Check the CAN bus

Detailed procedures and diagrams for checking CAN bus connections and identifying faults.

Checking the boards

Steps for visual inspection, voltage measurement, and replacement of circuit boards.

Device leakage current too high

Troubleshooting steps for high device leakage current, involving checking cable shields and replacing components.

Error analysis of X-RAY control signal path: up to unit hardware version AG (with board DX41)

Troubleshooting guide for X-RAY control signal path issues specific to units with board DX41.

Error analysis of X-RAY control signal path: unit HW version BA or higher (without board DX41)

Troubleshooting guide for X-RAY control signal path issues on units with HW version BA or higher.

Adjustment

Important information concerning adjustment

Crucial warnings and notes before performing system adjustment procedures.

Diaphragm/system adjustment menu

Explains how to access and navigate the Diaphragm/System Adjustment menu in SIDEXIS XG.

Adjusting the panoramic X-ray unit

Detailed procedures for adjusting the panoramic X-ray unit, including needle phantom usage.

Adjusting the cephalometer (XGPlus / 3Dready / XG 5, if ceph present)

Procedures for adjusting the cephalometer, including functional tests and specific settings.

TSA sensor adjustment (only for XGPlus with full version)

Guide for adjusting the TSA sensor, focusing on image evaluation and parameter settings.

Service routines

Service routines with SIDEXIS

Procedures for running service routines that require the SIDEXIS software.

Service routine S017

Comprehensive guide to various configuration settings, including hardware version, language, remote control, and PPE display.

Service routine S018

Procedures for height adjustment, including setting travel limits and testing sensors.

Service routine S034

Procedures for displaying and calibrating center positions of Ceph scan sensor axes and secondary diaphragm axes.

Repair

Safety check

Essential safety checks, including ground wire and leakage current tests, before and after repair.

Replacing the height adjustment motor (M1_4)/spindle

Detailed steps for preparing, moving the slide, and replacing the height adjustment motor/spindle.

Replacing the ring motor (M1_3)

Procedures for replacing the ring motor, including reusing coupling and installing the new motor.

Replacing the X-ray tube assembly

Critical procedure for replacing the X-ray tube assembly, emphasizing safety precautions and software updates.

Replacing circuit boards

Important notes and procedures for replacing circuit boards like DX1, DX11, DX32, DX41, DX42, DX89, DX91.

Maintenance

Checking the height adjustment

Procedures for checking height adjustment, including motor, rod, limit switches, and audible signals.

Checking the tube data

Procedures for checking kV actual value, radiation time, and tube current with specific service routines and test equipment.

Checking the protective ground wires

Detailed steps for measuring protective ground wire resistance and checking connections.

Checking the unit leakage current

Procedure for measuring unit leakage current using an automatic tester or manual methods.