65 45 490 D3286
12 D3286.201.03.15.23 04.2015
6Follow-up Sirona Dental Systems GmbH
6.1After each treatment session Operating Instructions SPRAYVIT / C8 / E / L / M
6
Follow-up
6.1
After each treatment session
Immediately after treatment
NOTICE! Condition immediately, or at the latest, one hour after treatment.
SPRAYVIT
✔ Wear appropriate protective clothing.
1. Remove any residue, e.g. from impression material or caustic
chemicals immediately.
2. Purge the water and air channels on the treatment center for 30
seconds.
3. Pre-disinfect directly at the treatment center [ → 13].
4. Detach the nozzle and the handle sleeve [ → 10].
5. Transport the nozzle and handle sleeve to the hygiene room in a
suitable transport container.
6. Conduct automatic conditioning [ → 13]. Manual conditioning [ → 13]
is possible in exceptional cases if the national/local regulations are
followed.
7. Sterilize the nozzle and handle sleeve and accessories [ → 14].