65 45 490 D3286
D3286.201.03.15.23 04.2015
13
Sirona Dental Systems GmbH 7Conditioning
Operating Instructions SPRAYVIT / C8 / E / L / M 7.1Conduct pre-disinfection
7
Conditioning
7.1
Conduct pre-disinfection
Predisinfection _ Pr oduct name _ Variable
✔ Wear appropriate protective clothing.
✔ All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.
1. Spray the surface with disinfectant.
2. Wipe the disinfectant away with a cloth.
For further processing, SPRAYVIT / C8 / E / L / M should be dry and
free of residue.
Disinfectants in USA und Canada
In the USA and Canada, for example, you can use:
●CAVICIDE
®
● CAVIWIPES
™
Please observe the manufacturer's instructions for using instrument
disinfectants.
7.2
Automated cleaning and disinfecting ...
Except valve body
NOTICE! The valve body is not suitable for conditioning.
7.2.1 ... with cleaning and disinfection equipment
Exception _ product name
NOTICE! SPRAYVIT / C8 / E / L / M is not suitable for use with a cleaning
and disinfection device.
7.3
Manual cleaning and disinfection
Immediately after treatment
NOTICE! Condition immediately, or at the latest, one hour after treatment.
Except valve body
NOTICE! The valve body is not suitable for conditioning.
NOTICE! Never clean in an ultrasound bath!
NOTICE! Never immerse in disinfectants!
Use soft brush
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
Procedure _ product name _ Brushing off, disinfection, packaging, sterilization
✔ Wear appropriate protective clothing.
✔ All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.
1. Brush SPRAYVIT / C8 / E / L / M under running water (< 38 °C,
< 100 °F, at least drinking water quality) and good lighting (min. 500
lux) and color rendering index (min. 80 Ra) until no more dirt can be
seen, for at least 10 seconds.
2. Conduct thermal disinfection or unwrapped steam sterilization.
3. Pack SPRAYVIT / C8 / E / L / M in packaging material suitable for
sterilization and storage. e.g. paper/plastic composite packaging.
4. Sterilize the adapter [ → 14].