65 45 490 D3286
8 D3286.201.03.15.23 04.2015
3Technical description Sirona Dental Systems GmbH
3.3Technical data Operating Instructions SPRAYVIT / C8 / E / L / M
Electronic supply (if required)
Media supply
Operating conditions
Year of manufacture On the valve body.
This product bears the CE marking in accordance with the provisions of
Council Directive 93/42/EEC of June 14, 1993 concerning medical
devices.
Water heating power in W 60 - 120 (max. 22 V)
Air heating power in W ~ 15 (max. 8 V)
Heating cartridge, electrical
resistance in Ω
4.2 ± 0.21
Lamp voltage in V 3.4 ± 0.1
LED current in mA 100 ± 5
SPRAYVIT SPRAYVIT C8 SPRAYVIT E SPRAYVIT L SPRAYVIT M
At the supply hose inlet, unit end:
Temperature of input medium
in °C
18 ± 3 max. 38 max. 38 18 ± 3 18 ± 3
Water pressure in bar
(flowing)
2 ± 0.1 2 ± 0.1 2 ± 0.1 2 ± 0.1 2 ± 0.1
Water flow rate in ml/min 130 130 130 130 130
Air pressure in bar (flowing) 3.2 ± 0.1 3.2 ± 0.1 3.2 ± 0.1 3.2 ± 0.1 3.2 ± 0.1
Air flow rate in Nl/min 13 13 13 13 13
Ambient temperature 10°C - 40°C, (50°F - 104°F)
Relative humidity in % 30 - 85
Air pressure in hPa 700 - 1060