EMC Information
Guidance and manufacturer's declaration -Electromagnetic
immunity
The U80H is intended for use in the electromagnetic environment
specied below. The customer or the user of the U80H should ensure
that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
Environment-
guidance
Electrostatic
Discharge(ESD)
IEC 61000-4-2
±6 kV contact
±8 k V air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic
tile. If oors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical test
Transient/burst
IEC 61000-4--4
±2 kV for
power
Supply lines
±1 kV for
input/output
lines
±2 kV for
power
Supply lines
±1 kV for
input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV
differential
mode
±1 kV
differential
mode
Mains power quality
should be that of a
typical
±2kV
common
mode
±2kV
common
mode
commercial or hospital
environment.
22