No.PS※※-OMV0006-D
Product individual parameters
Index
(dec)
Sub
index
Access
1
Parameter
Data
storage
2
Date
type
3
Initial value
(dec)
Remarks
0x03E8
(1000)
0 R/W
Unit
(Selection of
display unit)
Y U8
ISE20B-L(-M): 0
ISE20B-L-P: 4
ZSE20B-L(-M): 1
ZSE20B-L-P: 4
ZSE20BF-L(-M): 1
ZSE20BF-L-P: 4
Setting of display unit
0: MPa
1: kPa
2: kgf/cm
2
3: bar
4: psi
5: incHg (ZSE20B(F) only)
6: mmHg (ZSE20B(F) only)
0x03E9
(1001)
0 R/W
NorP
(Selection of
PNP/NPN)
Y U8
0x01
(1)
Setting of switch output specification
0: NPN
1: PNP
0x03EA
(1002)
0 R/W
CoL
(Selection of
display color)
Y U8
0x02
(2)
Setting of display color
0: red (Constantly red)
1: Grn (Constantly green)
2: 1SoG
(OUT1 turns green at ON)
3: 1Sor
(OUT1 turns red at ON)
4: 2SoG
(OUT2 turns green at ON)
5: 2Sor
(OUT2 turns green at ON)
0x03EB
(1003)
0 R/W
diAg
(Selection of
diagnostic
information)
Y U8
0x01
(1)
Sets the diagnostic information bit
of the input process data
0: Unused (Constantly OFF)
1: All errors ON
Product internal failure
Residual pressure error
Device temperature upper limit
exceeded
Measured pressure range
upper limit exceeded
Falls below measured pressure
range lower limit
2: When the following errors occur,
the parameter turns ON
Measured pressure range
upper limit exceeded
Falls below measured pressure
range lower limit
0x03F2
(1010)
0 R/W
oUt1
(Selection of
OUT1 output
mode)
Y U8
0x00
(0)
Setting of OUT1 output mode
0: HYS (Hysteresis)
1: Wind (Window comparator)
2: Err (Error output)
3: oFF