If dressing wound with gauze
Apply RENASYS So Port
Apply a layer of saline moistened gauze to wound bed.
Unfold remaining saline-moistened gauze and loosely ll
the entire wound cavity. Avoid over packing the wound.
Precaution: If multiple pieces ofwound filler are needed tofill
the wound profile, count andrecord how many pieces are
present to ensure all pieces are removed at a dressing change
to minimize the risk of retention andpossible infection.
Use of gauze as a filler in wounds that are in a weight bearing
location or have moderate toheavy drainage may challenge
the transfer of fluid andvacuum. Foam is recommended
inthese wounds.
While holding the transparent lm, expose one side
ofthe adhesive backing byremoving a single panel,
andapply over the wound.
Cover wound ller with transparent lm, removing
remaining adhesive panels toseal, then the top stabilization
panel. Film should securely anchor 5cm/1.97in beyond
wound margin to facilitate adequate seal.
Note: Avoid stretching or pulling the transparent film tominimize
tension ortrauma to the periwound skin. Overlap the edges
ofthe transparent film byaminimum of7.5cm/2.95in whenusing
multiple pieces of transparent film.
Remove the adhesive backing panel from the RENASYS
So Port dressing, and align the port opening directly
over the hole in the transparent lm.
Use gentle pressure to anchor the So Port tothe
transparent lm.
Cut a circular opening (no less than 2cm/0.75in in
diameter) in the center ofthelm, over thewound ller.
Remove any loose transparent lm anddispose of away
from the wound.
4b
6
5
7